Administration of Eleview®
Administration of Eleview®

Eleview® is designed to save time with streamlined preparation. No special equipment is needed. Sterile, premixed ampoules are ready to use.
Twist-and-go ampoules

Open

Twist

Go
Review the administration guide
Use this quick reference guide as a reminder of the simple administration steps for Eleview®. For added support, see the Instructions for Use sheet for Eleview®.
ASGE Tech Talk: Administration
View transcript

Fits into your practice
Eleview® saves preparation time for your staff, is a low-viscosity emulsion that is easy to inject, and works with a wide array of endoscopic injection needles.1,*
- Use with a standard, commercially available 23-gauge to 26-gauge needle1,*
- Ampoules are supplied with a female Luer-lock closure to enable extraction with a common syringe1
- Premixed with methylene blue1
- Reduces or eliminates compounding
- Sterile1
- An FDA-cleared medical device2
*Injection needle not included with Eleview®.

Preparation of Eleview® for injection is as follows:
- Open the aluminum pouch, detach one ampoule from the strip, and then open the ampoule by twisting off the cap following the arrow direction.
- Screw a disposable, sterile plastic or glass syringe with male Luer-lock connection fitting (not provided with the device) directly to the ampoule.
- Turn the ampoule upside down and extract the whole volume of the emulsion.
- Unscrew the disposable, sterile plastic or glass syringe from the ampoule and expel the air from the syringe until some drops of the emulsion pour out of the tip.
- Securely screw the filled Luer-lock syringe to a suitable sterile endoscopic injection needle (not provided with the device).
- Fill the endoscopic injection needle with the emulsion and expel any air.
- Infuse a small quantity of the emulsion to determine the amount of force needed to inject the material.
- Retract the needle.
- Introduce the endoscopic injection needle through the working channel of the endoscope.
- Inject Eleview® immediately beneath the lesion to be excised in one or in a series of injections around the base of the lesion.
- Discard the empty ampoule and store the unused ampoules inside the aluminum pouch in the folding box.